Case Report: Extending the Range of Vascular Access in Anatomically Challenging Patients With the GORE® Hybrid Vascular Graft

By Patricia Rosenberry, MS, BSN; Soo Yi, MD; Silke Niederhaus, MD; and David B. Leeser, MD, MBA
 

The GORE® Hybrid Vascular Graft (HVG) was approved for use in the United States in 20101 and was first used in 2011. One year later, the HVG with the 10 cm Nitinol Reinforced Section (NRS) was introduced in the United States. The device uses a deployable stent to achieve a percutaneous and sutureless venous anastomosis. This has provided many novel ways to maintain vascular access in some of the most challenging patients.1-4 The HVG with 10 cm NRS has further extended the reach of the access surgeon to a landing zone at the border of or even within the thorax.

We present two cases of patients with poor axillary veins of small caliber and previously failed upper extremity accesses who underwent successful dialysis graft placement using the HVG with 10 cm NRS with the tip of the nitinol-reinforced stent graft within the distal axillary/proximal subclavian vein.

CASE 1

A 73-year-old right-hand-dominant woman with a history of hypertension, diabetes mellitus, morbid obesity, and end-stage renal disease presented for access placement. The patient had previously failed peritoneal dialysis. In addition, the patient also had a previous attempt at a percutaneous and open hybrid graft placement by another surgeon. Vein mapping showed no suitable veins for fistula creation, but did show a patent axillary vein on the left upper extremity. The patient was therefore scheduled for an axillary venogram and possible hybrid graft placement.

The patient was taken to the operating room, and a venogram showed that the axillary vein was small in caliber and did not open into a suitable vein to support a graft until close to the thoracic wall (Figure 1). Because the suitable vein was so high in the axilla and close to the chest wall, a HVG with 5 cm NRS would not have reached the appropriate landing zone within the outflow vein. Therefore, a graft with 10 cm x 8 mm NRS was placed with a landing zone that was in a vein of a suitable caliber (Figure 2). The patient was seen 2 weeks postoperatively, and the graft had a palpable thrill and audible bruit. The graft was used shortly after the postoperative visit without difficulty.

Figure 1. Venogram showing a small-caliber vein (blue arrow) and larger vein near the thoracic cage margin (red arrow).

Figure 2. Deployed HVG.

CASE 2

A 41-year-old right-hand-dominant woman with a history of hypertension, diabetes mellitus, obesity, and end-stage renal disease presented for access placement. The patient had previously had a peritoneal dialysis catheter and a left upper extremity brachial-to-cephalic vein fistula that failed to mature. Because of the patient’s obesity, she was scheduled for a left arm venogram and hybrid graft placement.

The patient was taken to the operating room, and a venogram again showed a poor outflow vein that was small in caliber and opened into an appropriate vein near the chest wall (Figure 3). In this case, a HVG with 10 cm x 7 mm NRS was placed, again to reach a more proximal suitable landing zone for the end of the stent portion of the graft (Figure 4).

Figure 3. Venogram showing small-caliber vein (blue arrow) and larger vein near the thoracic cage margin (red arrow).

Figure 4. Deployed HVG.

The patient was seen postoperatively and had an audible bruit and a palpable thrill. The graft was used shortly after the postoperative visit.

DISCUSSION

The HVG is an important tool for all access surgeons to have in their toolbox. As seen in the previously described clinical scenarios, which support other published literature regarding the use of the HVG, this device allows for the preservation of upper arm access in patients with failing or failed upper arm grafts, morbid obesity, or previously stented venous outflow.4 This graft can avoid morbidity in obese patients (due to a smaller axillary incision) who present difficult access challenges with a secondary patency of over 60% at 1 year.5 The HVG allows for the creation of a sutureless anastomosis in cases where a standard graft could not be successfully placed. In these cases, the newer graft with the extended 10 cm stent portion allows for the creation of a graft with adequate outflow in a patient with a small-caliber or diseased axillary vein that does not open into a good outflow vein until very close to the thorax, where the vein becomes the subclavian.

Surgeons who develop the required skills to create access options using the HVG will become essential members of their access team and provide life-preserving care for patients with end-stage renal disease. As we all know, it is not the first access placed that is the most important; maintaining access in patients with decreasing options is the challenge. Access numbers four, five, six, seven, and beyond are the ones that sustain a patient’s life line. The HVG with 10 cm NRS has an important role to play in the ongoing access challenge.

1. US Food and Drug Administration. 510(k) summary of substantial equivalence: GORE HYBRID Vascular Graft. https://www.accessdata.fda.gov/cdrh_docs/pdf9/K093934.pdf. Accessed April 23, 2017.

2. Afaneh C, Aranova A, Ross JR, Leeser D. Use of hybrid vascular access grafts in failing access for hemodialysis: report of two cases. J Vasc Access. 2012;13:513-515.

3. Jones RG, Inston NG, Brown T. Arteriovenous fistula salvage utilizing a hybrid vascular graft. J Vasc Access. 2014;15:135-137.

4. Yi S, Rosenberry P, Leeser DB. Treating disadvantaged venous anatomy with the GORE® hybrid vascular graft. Endovasc Today. 2016;16:17-19.

5. Anaya-Ayala JE, Davies MG, El-Sayed HF, et al. Early experience with a novel hybrid vascular graft for hemodialysis access creation in patients with disadvantaged anatomy. J Endovasc Ther. 2015;22:778-785.

Patricia Rosenberry, MS, BSN
University of Maryland Shore Regional Health
Easton, Maryland
Disclosures: None.

Soo Yi, MD
Assistant Professor of Surgery
University of Maryland School of Medicine
Baltimore, Maryland
Disclosures: None.

Silke Niederhaus, MD
Assistant Professor of Surgery
University of Maryland School of Medicine
Baltimore, Maryland
Disclosures: None.

David B. Leeser, MD, MBA
Associate Professor of Surgery
University of Maryland School of Medicine
Baltimore, Maryland
dleeser@smail.umaryland.edu
Disclosures: None.

 

Contact Info

For advertising rates and opportunities, contact:
Craig McChesney
484-581-1816
cmcchesney@bmctoday.com

Stephen Hoerst
484-581-1817
shoerst@bmctoday.com

Charles Philip
484-581-1873
cphillip@bmctoday.com

About Endovascular Today

Endovascular Today is a publication dedicated to bringing you comprehensive coverage of all the latest technology, techniques, and developments in the endovascular field. Our Editorial Advisory Board is composed of the top endovascular specialists, including interventional cardiologists, interventional radiologists, vascular surgeons, neurologists, and vascular medicine practitioners, and our publication is read by an audience of more than 22,000 members of the endovascular community.