Endovascular Stent-Grafting in a Patient With a Rapidly Enlarging Thoracic Aortic Aneurysm

By Prashanth Vallabhajosyula, MD, MS, and Wilson Y. Szeto, MD
 

A 79-year-old woman with a history of Sjögren’s syndrome, rheumatoid arthritis, hypertension, moderate to severe interstitial lung disease with pulmonary fibrosis, and a known descending thoracic aortic (DTA) aneurysm (4.5 cm) presented with rapid expansion of the aneurysm to a maximal diameter of 6.2 X 5.6 cm over 1-year follow-up. Given these findings, the patient was electively taken for endovascular surgical intervention. The aneurysm had a large mural thrombus burden along with marked tortuosity in the distal thoracic aorta, extending from the proximal DTA to the aortic hiatus. The aortic dimensions were 29 mm at the left subclavian artery takeoff and 28 mm at the celiac axis.

Intraoperatively, right common femoral arterial access was achieved via open groin exposure, along with percutaneous left common femoral artery access for diagnostic study. Upon heparinization, a guidewire was placed into the ascending aorta under fluoroscopic guidance and exchanged to a Lunderquist stiff wire (Cook Medical, Bloomington, IN). A pigtail catheter was advanced from the left femoral access 7 F sheath, and angiography delineated the aortic arch/DTA anatomy (Figure 1). A 34 to 30 mm taper X 200 mm Relay graft (Bolton Medical, Inc., Sunrise, FL) was maneuvered around a severely angulated DTA by the aortic hiatus and was deployed proximally by the left subclavian artery without occluding it (Figure 2A). Upon confirming good proximal seal with angiography, DTA anatomy by the aortic hiatus was delineated, with special attention given to the celiac axis takeoff and the severe angulation of the aorta. Next, a second 34 X 34 X 150 mm Relay graft was advanced in a retrograde fashion and parked inside the first stent with a 5 cm overlap. The graft was then deployed, ensuring that the distal landing zone did not occlude the celiac axis (Figure 2B). A completion angiogram showed good proximal and distal seal, without any type I or III endoleaks (Figure 2C). The patient had an uneventful recovery and was discharged to home on postoperative day 6.

At her most recent 2-year follow-up, the patient continued to do well and remained asymptomatic. A CT scan at follow-up showed no endoleaks (Figure 3), with reduction of the DTA aneurysm to 5.2 cm.

The Bolton Relay graft was the ideal endograft choice in this case given the extreme tortuosity seen in the DTA by the diaphragmatic hiatus. The flexibility of this device, which was specifically designed to handle such tortuosity in the thoracic aorta, facilitated excellent coverage of the DTA aneurysm without any endoleak. The modular design of the system, along with multiple straight and tapered configurations and different lengths, enabled ease of coverage of the entire DTA aneurysm with two stent-graft segments. Proximal landing with a 34 X 30 mm tapered device enabled good proximal landing seal. The ability to place a 34 X 34 mm distal stent-graft piece with 5 cm stent overlap allowed excellent seal between the two stent-grafts to prevent type III endoleak and a good distal landing zone seal by the celiac axis (DTA measured 28 mm by the aortic hiatus). The S-bar technology of the Relay system enables conformability to handle severe angulation of the aorta and provides longitudinal support to prevent the stent-graft from collapsing at the site of a severely kinked aorta. In this case, the severe angulation of the DTA aneurysm by the diaphragmatic hiatus was handled with relative ease due to the excellent conformability of the Bolton Relay endoprosthesis.

Prashanth Vallabhajosyula, MD, MS, is Assistant Professor of Surgery, Division of Cardiovascular Surgery, University of Pennsylvania Medical Center in Philadelphia, Pennsylvania. He has disclosed that he has no financial interests related to this article.

Wilson Y. Szeto, MD, is Associate Professor of Surgery; Surgical Director, Transcatheter Cardio Aortic Therapies; Associate Director, Thoracic Aortic Surgery, Division of Cardiovascular Surgery, University of Pennsylvania Medical Center in Philadelphia, Pennsylvania. He has disclosed that he was the Principal Investigator at the University of Pennsylvania for the Bolton Relay trial. Dr. Szeto may be reached at wilson.szeto@uphs.upenn.edu.

 

Contact Info

For advertising rates and opportunities, contact:
Craig McChesney
484-581-1816
cmcchesney@bmctoday.com

Stephen Hoerst
484-581-1817
shoerst@bmctoday.com

Charles Philip
484-581-1873
cphillip@bmctoday.com

About Endovascular Today

Endovascular Today is a publication dedicated to bringing you comprehensive coverage of all the latest technology, techniques, and developments in the endovascular field. Our Editorial Advisory Board is composed of the top endovascular specialists, including interventional cardiologists, interventional radiologists, vascular surgeons, neurologists, and vascular medicine practitioners, and our publication is read by an audience of more than 22,000 members of the endovascular community.