FDA Issues Safety Alert on MRA Follow-Up of Neurovascular Embolization Coil Procedures

March 12, 2018—The US Food and Drug Administration (FDA) has issued a Safety Alert providing information about the potential for increased image artifact associated with MRA imaging for patient follow-up of certain procedures following neurovascular embolization coiling. The letter to health care providers is available online here.

According to the agency, neurovascular embolization coils are used in the treatment of brain aneurysms and other vascular abnormalities. Digital subtraction angiography (DSA) has traditionally been the method used for imaging aneurysms after coiling. However, with advances in MRA image quality and accessibility, many health care providers may choose to use MRA in lieu of DSA because patients are not exposed to ionizing radiation or embolic stroke risks.

Information reviewed by the FDA, including the FDA's medical device reports (MDRs), information from device manufacturers, and published literature, indicates that when MRA is performed on patients implanted with neurovascular embolization coils containing 304V stainless steel (either as part of the coil implant itself or left behind as part of the detachment process), the images may contain larger than expected MRA artifact or image voids when compared with other metals. In these cases, the reduced quality of the MRA image from increased artifact can result in inaccurate clinical diagnoses (eg, occlusion status) and subsequent inappropriate medical decisions.

The FDA will continue to monitor this situation and will update this communication if significant new information becomes available.

The Safety Alert advised that most neurovascular embolization coils currently on the market are labeled as "magnetic resonance (MR) conditional." The MR conditional information in the product labeling conveys to the user the conditions of safe use of the device when scanned within an MR environment. The product labeling for neurovascular embolization coils does not currently convey the extent of MRA image artifact caused by the device or the MRA imaging parameters that will yield the lowest amount of image artifact if MRA is used for patient follow-up.

The FDA recommends that health care providers:

  • Be aware of the presence of 304V stainless steel in the coil system(s) typically used (either as part of the coil implant itself or left behind as part of the detachment process). If uncertain, the applicable device manufacturer should be contacted for information about the coil, its detachment mechanism, and any specific recommendations regarding the use of MRA with their product.
  • Understand that 304V stainless steel–containing embolization coil systems may increase image artifact on MRA exams performed at the time of aneurysm follow-up. In these situations, x-ray–based DSA should be considered.
  • If choosing to use MRA for follow-up, utilize optimal imaging parameters to minimize image artifact, including shortest echo times and high readout bandwidth. In addition, ensure that the MRI system at your site meets all conditions provided in the MRI conditional labeling of the coil system (eg, magnetic field strength in units of Tesla).
  • Report any event of image artifact associated with MRA that might have led to misinterpretation of the image to MedWatch, the FDA Safety Information and Adverse Event Reporting Program online voluntary reporting form. The report should include information on the manufacturer(s), brand(s) of coils used, imaging parameters utilized, and follow-up action taken (eg, DSA completed, DSA results, and applicable interventions).
  • Device manufacturers and user facilities must comply with the applicable MDR regulations. Health care personnel employed by facilities that are subject to FDA's user facility reporting requirements should follow the reporting procedures established by their facilities. Prompt reporting of adverse events can help the FDA identify and better understand the risks associated with medical devices.

Health care professionals and patients are encouraged to report adverse events or side effects related to the use of these products to the FDA's MedWatch Safety Information and Adverse Event Reporting Program at www.fda.gov/MedWatch/report.