Placement Accuracy and Customization

Prof. Jose M. Abadal, MD, PhD; Prof. Miguel Araujo, MD; and Prof. Esther Vazquez, MD, discuss their clinical experience using this innovative stent graft system.

By Prof. Jose M. Abadal, MD, PhD; Prof. Miguel Araujo, MD; and Prof. Esther Vazquez, MD
 

Prof. Jose M. Abadal, MD, PhD
Department of Vascular and Interventional Radiology
Severo Ochoa University Hospital
Madrid, Spain
jmabadal@yahoo.es
Financial disclosures: None.

Prof. Miguel Araujo, MD
Department of Vascular and Endovascular Surgery
Severo Ochoa University Hospital
Madrid, Spain
miguelaraujopazos@gmail.com
Financial disclosures: None.

Prof. Esther Vazquez, MD
Department of Vascular and Endovascular Surgery
Severo Ochoa University Hospital
Madrid, Spain
esther.vazquezro@salud.madrid.org
Financial disclosures: None.

An essential feature of any endovascular device used to treat abdominal aortic aneurysms (AAAs) is the ability to accommodate any necessary changes after the delivery system has been introduced. The INCRAFT® AAA Stent Graft System (Cordis Corporation) was designed to solve the limitations of previous-generation endovascular devices, particularly with regard to versatility, accuracy, and real-time customization.

Profs. Abadal, Araujo, and Vazquez discuss two key attributes of the INCRAFT® AAA Stent Graft System—placement accuracy and in situ customization—and present several cases illustrating the device’s benefits in real-world clinical practice.

Why did you decide to first try the INCRAFT® AAA Stent Graft System?

We have extensive experience in AAA endovascular treatment, but in working with other endografts over the years, we have encountered problems, such as difficulties with femoral and iliac access and navigation as well as a need for multiple device sizes and lengths to account for different anatomies. The launch of the INCRAFT® System has made our daily practice easier, not only because of the ultra-low profile and flexible delivery system, but also the trimodular design of the graft, which allows the procedure to be optimized using the fewest prostheses and covers a broad number of cases with simplified inventory management.

How have you adopted the INCRAFT® AAA Stent Graft System in your practice?

The INCRAFT® System has become a device that is routinely used in our practice. Specifically, it is our first choice to treat AAA patients with narrowed and diseased aortoiliac vessels and access. The main benefits of the INCRAFT® AAA Stent Graft System are the device’s ultra-low profile and its in situ iliac extension customization.

Why is placement accuracy important to you?

Placement accuracy is one of the key goals of the AAA endovascular procedure and enables a great result, reducing any unexpected complications. Placement accuracy is also related to long-term clinical success. Device misplacement can result in endoleaks and persistence of aneurysmal sac enlargement, which increases endovascular and surgical complexity and risk of reintervention.

In your experience, how does the placement accuracy of the INCRAFT® System compare to other devices?

Placement accuracy of the INCRAFT® System is enhanced by the distinctive proximal and distal radio­paque markers that can be partially repositioned prior to full deployment. Although every case is planned in advance with precise measurements, and the optimal graft is selected for each patient, there is always a minimal intraprocedural discrepancy. The INCRAFT® device is versatile and aids in intraprocedural graft adjustment, without requiring other unplanned grafts. That is an important difference from competitors and a very important issue in emergency cases.

What are the benefits of placement accuracy of the INCRAFT® System to the patient, physician, and institution?

The INCRAFT® endograft offers customized treatment to a broader range of patients. Moreover, it may allow patients with AAAs to be treated who otherwise would not be candidates for endovascular repair using other devices because of anatomic constraints, for example, narrowed iliac vessels (diameter < 7 mm) or tight aortoiliac bifurcation (10-mm limb diameter).

For the clinician, the easier the procedure, the better results in terms of surgery, access and placement complications, grafts needed, radiation exposure, etc. Our feeling is that the anatomic variations during the surgery do not have an important impact on your planning and resource management.

For the institution, we can stock less inventory using the INCRAFT® System compared with any other endoprostheses, because of its in situ length customization. In our experience, the flexibility of the INCRAFT® System allows placement in up to 90% of AAA procedures. Billing prediction per procedure has been simplified because fewer product codes are needed, and there is a low variation in the number of devices used in patients because of its trimodular design.

Does improved placement accuracy lead to a reduction in acute and chronic complications?

Data on placement accuracy from the INNOVATION study have demonstrated a 2-mm median distance from the lowest renal artery to the graft edge markers in 58 interventions, with no reports of stent migration in 50 cases at 2 years. Data on in situ limb adjustment also outline distal limb accuracy with a median 12-mm distance from the origin of the internal iliac artery. These technical data correlate with the excellent 1-year results and absence of complications: 0% type I or III endoleaks and no aneurysmal sac enlargement.1

Can you explain in situ length customization? How does this work? What are the benefits of this feature?

In situ length customization allows a clinician to adjust the graft in real time during the procedure. The limb length can be adjusted bilaterally, up to 3 cm ipsilaterally and 2 cm contralaterally. This substantially improves placement accuracy and reduces the risk of inadvertent side branch coverage.

The safety of variable limb overlapping is enhanced by suture knots in the outer surface of the limb stent, which provides more stability and firms up the modular junction.

The benefit is obvious, because you can use the same iliac extension for different lengths; as a result, a broad range of patients can be treated with a small stock. This issue is very important in ruptured AAAs. The device’s versatility allows us to treat patients easily and quickly, even in emergency cases, when there is little time for planning and measurements.

What are the main benefits of the INCRAFT® System as compared to other devices you use in your practice?

First, the iliac limbs of the main graft have 11 mm in diameter; this feature is important in small and diseased aortic bifurcations. The very low profile system (equivalent to a 12-F catheter sheath introducer profile*) allows the device to be introduced through the superficial femoral artery in high and diseased femoral bifurcations. The delivery system is extremely easy and precise for suprarenal fixation and the aortic main body, and without a cap at the top, deployment and the retrieval of the graft is simplified. The learning curve is also reduced when compared to other devices. 

*The iliac limb delivery system has a 12-F outer diameter for prothesis diameters between 10 mm and 20 mm and the 13-F outer diameter for the 24-mm diameter prosthesis.

1. Torsello G, Scheinert D, Brunkwall JS, et al. Safety and effectiveness of the INCRAFT AAA Stent Graft for endovascular repair of abdominal aortic aneurysms. J Vasc Surg. 2015;61:1-8.

CASE 1: NARROWED AORTIC BIFURCATION

An 80-year-old man with multiple risk factors, including hypertension, hyperlipidemia, and coronary heart disease, presented with a 60-mm AAA. CT showed a narrowed distal aorta (Figure 1). An angiogram demonstrated a narrowed distal aorta with nontortuous iliac arteries. The INCRAFT® device with a 22-mm main body device and 10-mm diameter iliac limbs was deployed (Figure 2). Kissing balloon angioplasty was performed to maintain patency without the need for stent reinforcement. AAA exclusion was successful without limb kinking or occlusion on follow-up.

The INCRAFT® device may broaden the inclusion criteria in narrowed iliac bifurcation due to the ultra-low profile and iliac limbs of 10-mm in diameter. The use of an aorto-uni-iliac graft and femorofemoral bypass may also be avoided. Iliac graft radial force was sufficient to maintain patent bifurcation.


CASE 2: TORTUOUS AORTOILIAC ARTERIES

A 73-year-old man presented with a 68-mm diameter AAA and renal insufficiency. CT showed tortuous and calcified iliac vessels and a long neck infrarenal AAA (Figure 1). Percutaneous endovascular aortic repair with the INCRAFT® device (22-mm main body) was performed. The device easily handled the patient’s challenging anatomy. There was no need to force the stent graft delivery system, and a catheter-like navigation sensation was felt. There was no excessive rectification of the iliac vessels/aorta (Figure 2). The INCRAFT® AAA Stent Graft adapted to the anatomy successfully, excluding the aneurysm (Figure 3).

The INCRAFT® AAA endovascular device allowed a catheter-like, easy navigation, even in hostile iliofemoral vessels. The device design reduces trauma to the iliac artery access, avoiding potential complications (eg, rupture, dissection/thrombosis, embolism).

Exercise particular care in areas that are difficult to navigate, such as areas of stenosis, intravascular thrombus, calcification or tortuosity, or where excessive resistance is experienced, as vessel or catheter damage could occur. Consider performing balloon angioplasty at the site of a narrowed or stenotic vessel, and then attempt to gently reintroduce the catheter delivery system. Also exercise care with device selection and correct placement/positioning of the device in the presence of anatomically challenging situations such as areas of significant stenosis, intravascular thrombus, calcification, tortuosity, and/or angulation, which can affect successful initial treatment of the aneurysm.


CASE 3: NARROWED ILIAC VESSELS

A 75-year-old patient presented with an asymptomatic AAA of 60 mm in diameter. The patient had comorbid factors that increased surgical risk, including multiple drug allergies, renal failure, previous abdominal surgeries, and cardiovascular disease, as well as a vascular history of Rutherford stage 2 peripheral artery disease with bilateral femoropopliteal occlusion.

CT of the anatomy was favorable, as the aneurysm had a long infrarenal neck and no angulations. However, the iliac arteries were extensively diseased and narrowed in diameter (Figure 1). The peripheral arteries were also narrowed, and there was a bilateral femoropopliteal occlusion. At this point, the main concern was to choose a graft that could manage the extensively diseased iliac arteries and avoid graft component overlapping.

We decided to use the INCRAFT® AAA Stent Graft System because the ultra-low profile would facilitate iliac navigation through the tortuous, stenotic, and narrowed iliac vessels. Periprocedural customization of the iliac limbs would help enable accurate placement on a nondiseased landing zone. The main body graft was introduced and deployed through the right side, with an excellent navigation due to the ultra-low profile and catheter-like shaft flexibility. The contralateral left iliac stenosis was easily passed without the need of predilation because of the 12-F integrated delivery system.* In this case, it was mandatory to land the iliac extensions in the desired location to avoid any outflow problems (Figure 2). This is an advantage of the INCRAFT® device’s limb flexibility, which allows for a 2- to 3-cm in situ adjustment.

At 6-month follow-up, contrast-enhanced ultrasound demonstrated a reduction of the aortic sac (56 mm), without leaks, and the patient’s walking distance had improved as a result of the procedure (Figure 3).

Because of the combination of ultra-low profile and ease of navigation, the INCRAFT® AAA Stent Graft System may have an advantage over other devices in iliac vessels < 7 mm. This situation is more prevalent in women, Caucasian men of small stature, and Asian descendants.

*The iliac limb delivery system has a 12-F outer diameter for prothesis diameters between 10 mm and 20 mm and the 13-F outer diameter for the 24-mm diameter prosthesis.

Exercise particular care in areas that are difficult to navigate, such as areas of stenosis, intravascular thrombus, calcification, or tortuosity, or where excessive resistance is experienced, as vessel or catheter damage could occur. Consider performing balloon angioplasty at the site of a narrowed or stenotic vessel, and then attempt to gently reintroduce the catheter delivery system. Also exercise care with device selection and correct placement/positioning of the device in the presence of anatomically challenging situations such as areas of significant stenosis, intravascular thrombus, calcification, tortuosity, and/or angulation, which can affect successful initial treatment of the aneurysm.


CASE 4: ISOLATED COMMON ILIAC ANEURYSM

A 73-year-old man presented with an isolated aneurysm of the right common iliac artery. CT showed a 35-mm aneurysm in the common iliac artery that had grown over the 2 last years (Figure 1). The aneurysm started at the short neck of proximal iliac artery and ended at the level of the internal iliac bifurcation. There was an important discrepancy between the proximal common iliac and external iliac diameter.

Percutaneous access with two Perclose ProGlide devices (Abbott Vascular) was performed, and a 13/10-mm X 80-mm INCRAFT® limb graft was used to exclude the aneurysm. A 14-mm Amplatzer vascular plug (St. Jude Medical) was placed in the origin of the hypogastric artery to prevent retrograde flow into the aneurysm.

Care was taken to spare the gluteal and hypogastric branches and to preserve pelvic blood flow from the contralateral artery.

The endograft was successfully advanced and deployed precisely to the intended position. Sizing of the limb graft adapted to proximal and distal iliac diameters with the use of only one stent. Radial force of the stent graft and the interlocking suture knot design resulted in an adequate fixation of the stent graft without migration or endoleak.

Follow-up CT scan at 6 months demonstrated exclusion of the aneurysm, no stent graft migration, and preservation of hypogastric arterial branches (Figure 2).

The third-party trademarks used herein are trademarks of their respective owners.

 

Contact Info

For advertising rates and opportunities, contact:
Craig McChesney
484-581-1816
cmcchesney@bmctoday.com

Stephen Hoerst
484-581-1817
shoerst@bmctoday.com

Charles Philip
484-581-1873
cphillip@bmctoday.com

About Endovascular Today

Endovascular Today is a publication dedicated to bringing you comprehensive coverage of all the latest technology, techniques, and developments in the endovascular field. Our Editorial Advisory Board is composed of the top endovascular specialists, including interventional cardiologists, interventional radiologists, vascular surgeons, neurologists, and vascular medicine practitioners, and our publication is read by an audience of more than 22,000 members of the endovascular community.