Before You Place That Filter …
A guide to IVC filter placement and troubleshooting procedural challenges.
By Ulku Cenk Turba, MD; Saher S. Sabri, MD; Wael E.A. Saad, MD; Auh Whan Park, MD; John F. Angle, MD; and Alan H. Matsumoto, MD
Evidence-based guidelines from the American College of Chest Physicians and Society of Interventional Radiology recommend inferior vena cava (IVC) filter placement in patients with proven venous thromboembolism with a failure of, a contraindication to, or a complication secondary to anticoagulation.1,2 However, with the recent availability of retrievable or so-called optional IVC filters, the use of these devices has expanded significantly in patients who are perceived to be at high risk for pulmonary embolism (PE).1 In this article, we focus on the anatomical factors, procedural challenges, and complications related to the use of these devices.
PREPROCEDURAL EVALUATION
Assuming that there is a proper indication for IVC filter
placement, the procedure is performed using sterile
technique. In some instances, IVC filter placement has
been performed using intravascular ultrasound for
guidance.3 In most instances, an IVC gram is performed
at the time of the filter placement procedure. The IVC
gram is performed to define: (1) the IVC anatomy and
make sure that there are no anatomic variants such as a
duplicated IVC or a circumaortic or retroaortic renal
vein; (2) the diameter of the IVC (to make sure that the
IVC is not too large for the filter); (3) that there is no
IVC thrombus that would obviate or complicate the
placement of the filter; (4) the location of the renal
veins (because the optimum filter location is just below
the level of the lowest renal vein, so if thrombosis of the
IVC filter occurs, there is minimum dead space above
the filter for thrombus to form and serve as a source of
significant PE); and (5) that the catheter is actually in the IVC and not in a parallel vein such as a lumbar vein.
If a cross-sectional imaging study such as a computed
tomography (CT) or magnetic resonance imaging (MRI)
scan is available and defines the IVC and renal venous
anatomy, the IVC gram is obtained mainly for the purposes
of items 3 to 5 mentioned previously. The CT
and/or MRI are also reviewed to make sure there are no
extrinsic masses that are compressing the IVC.
A marker or calibrated multisidehole catheter should be used and positioned in the field of view to provide an internal calibration reference. Placing the catheter at the confluence of the IVC and the common iliac veins and power injecting 20 mL of contrast per second for 2 seconds results in reflux of contrast into both iliac veins to ensure that there is no IVC duplication, to fully define the IVC anatomy, and to demonstrate the “washout defects” due to the inflow of unopacified blood from the renal veins. Accurate measurements of the IVC length below the lowest renal vein and its diameter are determined based on the reference calibration catheter. After obtaining the IVC gram, the patient, the image intensifier, and the procedure table should not be moved in order to minimize changes in parallax and to preserve the relationship between the venous anatomy and the bony landmarks (Figure 1). On occasion, it may be necessary to perform selective venography when an inflow defect suggests a circumaortic left renal vein (Figure 2) versus a prominent lumbar vein.
PERMANENT VERSUS OPTIONAL
IVC FILTERS
The primary advantage of using an optional filter
over a permanent filter is that an optional filter has the
possibility of being removed from the patient when it is
no longer needed or it can be left in place as a permanent
filter. Indeed, some physicians would argue that
there is really no future for permanent filters. Placement
of an optional filter has the additional advantage of
being able to be immediately retrieved and redeployed
if the initial deployment was not optimum (Figure 3).
However, there are no data that demonstrate the in
vivo durability of all of the currently available optional
IVC filters for > 5 years, and the retrievability of each
optional filter over time is not the same.
All currently available optional filters can be safely deployed in an IVC up to 28 to 32 mm in diameter, depending on the specific optional filter manufacturer indications for use. An IVC that is > 28 to 32 mm in diameter is called a megacava. In addition, the IVC is very elastic, and its diameter can vary depending on the volume status of the patient and whether the patient is performing a Valsalva maneuver at the time of the venogram. However, most IVCs are elliptical in shape. Therefore, the anterior-posterior diameter of the IVC tends to be less than the medial-lateral diameter, which provides some component of safety while determining the IVC diameter based on a venogram. When a megacava is encountered, there are a few available treatment options. The Gianturco-Roehm Bird's Nest filter (Cook Medical, Bloomington, IN), which is a permanent filter, can be placed in an IVC up to 40 mm in diameter. The Bird's Nest filter consists of four stainless steel wires (25 cm X 0.18 mm) attached to two V-shaped struts and is best suited for an IVC that is at least 7 cm in length below the renal veins. Alternatively, bilateral common iliac venous filters may be placed, especially when the use of an optional filter is desired.
LOCATION OF FILTER PLACEMENT
Most filters are placed in an infrarenal location,
because 90% of clinically significant PE originates from
the lower extremity or pelvic veins. Therefore, the optimal
position of an IVC filter is immediately below (for
nonconical filters) or just at the level of the lowest renal
vein (for conical filters) for the reasons previously
described. The inflow of blood from the renal veins will
tend to minimize clot formation craniad to an appropriately
placed infrarenal IVC filter. However, placement
of filters in the suprarenal IVC, bilateral iliac veins, superior
vena cava (SVC), or in both components of a duplicated
IVC (Figure 4) has been performed based on the
patient's anatomy and the likely source of PE.
Suprarenal placement of an IVC filter is performed if there is a duplicated IVC, if there is a large volume of thrombus within the infrarenal IVC or extrinsic compression of the IVC preventing the safe positioning of an infrarenal filter, or if there is extensive renal or gonadal venous thrombosis. There are some physicians that will place a filter in a suprarenal location in women who are likely to become pregnant. Filters placed in a suprarenal location are potentially subjected to chronic trauma from the overlying liver and may be more prone to fracture or penetrating the IVC over time. In addition, conical-shaped filters can severely tilt if their legs prolapsed into a hepatic vein. We experienced one patient who developed significant pain with deep inspiration after placement of a suprarenal filter and, upon inspection, we suspected that one of its legs was protruding into a hepatic vein and irritating the liver capsule during respiration. Upon removal of the filter, the pain immediately resolved.
Bilateral iliac vein filters are typically used in patients with a megacava (> 28–32 mm) or a duplicated IVC, or in patients with a retroaortic left renal vein component that drains into the IVC close to the iliac venous confluence.
Upper extremity and large volume internal jugular venous thromboses are being encountered with increasing frequency as the use of central lines increases. Due to the relatively smaller size of upper extremity veins, the risk for clinically significant PE is lower as compared to the risk associated with lower extremity and pelvic venous and IVC thrombosis. However, it may be a very challenging clinical situation when a patient with significant acute subclavian, axillary, and brachial venous thrombus presents with PE (and no other obvious source for the PE) and cannot be anticoagulated. In these types of cases, the benefits of placing an SVC filter must be weighed against the potential risks. The SVC is much shorter in length than the IVC, and the legs of a conical filter could prolapse into the azygous vein, causing the filter to tilt significantly. In addition, many of these patients undergo multiple central venous line placements, and the various wires and catheters could become engaged with the SVC filter and displace the filter. Constant cardiac motion and mediastinal pulsation could also cause the hooks or legs of the filter to penetrate the SVC, leading to a hemopericardium and tamponade. 4 Therefore, placement of an SVC filter should be only undertaken after careful consideration of each individual case, factoring in the anatomical challenges and the lack of good outcomes data with SVC filters.5 If a decision is made to place an SVC filter, the ideal location is one in which the filter legs are craniad to the azygous vein (which is usually outside of the pericardial reflection) and just caudal to the confluence of the brachiocephalic veins.
Choosing the Venous Access Site for Filter Placement
The right internal jugular or right common femoral
veins are the most common access sites for filter
deployment.6 With the availability of more flexible and
lower-profile filter devices, the left internal jugular, both
external jugular, left femoral, and upper extremity veins,
as well as a transcaval approach, have all been used as
access sites for IVC filter placement for a variety of reasons
(eg, extensive trauma, diffuse thrombosis, etc.).
The transcaval approach is the least often used
approach due to its complexity, but the left common
femoral vein approach is the least desired of the other
access site options for placement of a non-self-centering
conical-shaped filter. The acute angle of the left iliac
vein with the IVC directs the filter delivery sheath
against the right lateral aspect of the IVC wall, causing
the filter to be deployed in a tilted, and therefore less
functional, orientation (Figure 3).
Anatomic Considerations
The reported prevalence of a duplicated IVC is about
0.2% to 0.3%.7 With a duplicated IVC, the left IVC most
often drains into the left renal vein (Figure 4). There are
several reports of clinically
significant PE after placement
of an IVC filter in
patients with an undiagnosed
duplication of the
IVC.8 While obtaining an
IVC gram with a catheter
placed in the distal rightsided
IVC, it is important to
have contrast reflux well
into the left common iliac
vein; otherwise, a duplicated
IVC may be missed. On
occasion, a separate access
into the left common
femoral vein is needed,
because the iliac veins may
not join together but rather
empty into their respective
IVC moieties. Venographic
evidence that might alert
one to the possibility of a
duplicated IVC is a diminutive
right-sided IVC or a large amount of unopacified blood flowing from the
left-sided IVC into the left renal vein and subsequently
into the right-sided IVC. When these findings are
observed and there are no previous cross-sectional studies
that have excluded the presence of a duplicated IVC,
a left common iliac venogram should also be obtained.
When the presence of a duplicated IVC is diagnosed, the
filter can be placed in a suprarenal location or within
each IVC component.7 Our preference is to place a single
filter within the suprarenal IVC. Suprarenal IVC filters are
reported to be safe and effective in the prevention of PE,
despite some concern about the increased risk for filter
fractures.9
FILTER AND IVC THROMBOSES
The reported frequency of IVC thromboses varies
widely (0%–28%), depending on the consistency, duration,
and method of follow-up evaluation.10
Approximately 50% of IVC thromboses may be present
without patients reporting symptoms. Therefore,
detailed imaging of the IVC should be performed before
any filter retrieval attempt. Having stated this, our experience
with retrieval of Günther Tulip or Celect filters
(Cook Medical) at a mean of 43 days (1–343 days)
showed a very low incidence of clot burden in these filters
during venography at the time of their retrieval.6
However, our experience may be biased toward patients
that are less likely to have thrombophilia and therefore less likely to have clot in the filter at the time of its
retrieval. Chronic IVC occlusion/thrombosis in the presence
of a filter in situ is not uncommon and is probably
due to the specific patient populations receiving IVC filters
as well as the chronicity of the device. Systemic anticoagulation
is the first-line treatment in symptomatic
patients with IVC thrombosis and a filter. In symptomatic
patients, when anticoagulation fails, if the patient
has bulky acute/subacute thrombus, thrombolytic
agents may be used. If there is a chronically occluded,
partially recanalized, string-like IVC present, balloon
angioplasty and stent placement could be considered to
restore patency through the filter (Figure 5).
FILTER MISPLACEMENT
IVC filter misplacements outside of the confines of the
IVC have been reported, including placement of a filter
within the aorta11 and within the spinal canal.12 Mildly
or severely maldeployed filters (mild-to-severe tilting of
the filter or placement of the filter too caudal below the
renal veins or significantly covering the renal veins) are
not uncommon. If a misplaced IVC filter is a permanent
IVC filter, retrieval of the filter is much more challenging
or sometimes impossible, whereas if an optional filter is
used and maldeployed, it may be retrieved and repositioned
relatively easily (Figure 3).
FILTER TILT
What degree of tilting of a conical-shaped IVC filter
should be considered significant is debatable. However,
once tilting of a conical-shaped filter occurs, there are in
vitro data that show that the filtration efficiency of the
filter may be reduced. In a few published series, there is a
suggestion that conical-shaped filters with > 15° of tilting
from the longitudinal axis of the IVC may reduce its
filtration capability. Tilting of conical IVC filters has been
associated with recurrent pulmonary emboli.13,14
Therefore, if tilting occurs during deployment, filter
repositioning may be considered. Using optional filters
has significant advantage over permanent filters. Filter
retrieval and repositioning is possible even when the filter
is severely (> 45°) tilted.15
FILTER FRACTURE/MIGRATION
Filter fractures are not a common complication and
have been reported in < 1% of cases; however, pregnant
patients may be at greater risk for IVC filter fractures
due to extrinsic compression from a gravid uterus.16 When placing a filter in women of childbearing age, possible
long-term outcomes should be discussed with the
patient and the referring physician, and the indications
for filter placement should be carefully assessed.
Filter migration as a whole is also a rare complication. 17,18 It is important to keep in mind that fracture and migration of an IVC filter or recurrent PE should be suspected in a patient presenting with sudden, unexplained cardiopulmonary compromise and a known IVC filter in place. Although it is a rare complication, IVC filters can migrate to the heart and potentially lead to death. In most of these cases, the filter placed was not sized appropriately for the diameter of the IVC, the hooks of the filter did not adequately engage the caval wall, and/or a large thrombus caused the filter to become dislodged. Early detection of a migrated IVC filter may save a life. Once detected, a migrated filter may be retrieved endovascularly or surgically, depending on its location and its relationship with the cardiac structures (valves, chordae tendineae, papillary muscles, etc.). Filters can also become dislodged when intravascular procedures are performed and devices or guidewires become entangled with the filter. Most filters become fairly well endothelialized within 6 weeks, so a fair amount of force is needed to displace an appropriately sized and placed filter if it has been in place for a few months.
Filter Hooks and Strut/Leg Caval Penetration
Asymptomatic penetration of a filter hook and/or
strut/leg through the IVC wall is not uncommon.6
Probably the most important factor to assess is how
deep the hook or strut/leg has penetrated. Is the penetrating
hook and strut/leg at risk for injuring or penetrating
an adjacent organ such as the aorta, duodenum
(or any bowel), vertebral body, or ureter? Fortunately,
most patients with this problem are asymptomatic.
However, some patients may develop pseudoaneurysms,
infections, bowel injuries, or pain. If the filter is retrievable
and the patient is symptomatic, retrieval of the filter
should be performed. It may be necessary to place
another filter if the patient remains at risk for PE. If the
patient is asymptomatic, retrieval can be considered
depending on the estimated risk of doing nothing. In
some instances, surgical removal of the filter may be
necessary.
Fractures of the filter legs/struts have been more problematic with older versions of nitinol-based filters. Despite the many benefits of nitinol technology and its thermal memory characteristics, nitinol material tends to be more prone to fracturing.19 Even more recent designs have led to numerous case reports of fractured struts/legs, with the free fragments migrating into hepatic veins and pulmonary artery branches; however, there is an effort to change the structure of newer nitinol filters. Because the fragments are small and often difficult to see, they become difficult to retrieve. Whether or not these fragments may lead to future problems is not clear; however, microscopic evaluation of the retrieved filter and limbs revealed bending metal fatigue at the fracture sites. Percutaneous retrieval of filters with arm fracture or arm migration is recommended.20
FILTER RETRIEVAL
IVC filter retrieval success depends on filter dwell
time, the amount of filter hook and strut/leg penetration/
endothelialization, the amount of metal material
in contact with the IVC wall, and accessibility of the
retrieval hook on the filter to the retrieval device. A
variety of techniques and devices have been used to
maximize the filter retrieval success rate.6 In our institution,
the retrieval success rate for the Günther Tulip filter
is approximately 95% without significant complications,
even with filters in place for > 6 months. After
filter retrieval, a repeat IVC gram is performed to evaluate
the integrity of the IVC. In our reported experience,
there were no IVC stenoses > 40% and no extravasation.
6
In our unpublished experience of more than 350 optional filter retrievals (seven different optional filter types), < 1% of retrieval patients have undergone balloon dilation for IVC stenoses > 70%. Several patients who had stenoses between 40% and 70% were managed conservatively and underwent repeat venography or cross-sectional imaging at follow-up. These patients were asymptomatic and shown to have improvement in their IVC stenoses. Although we have not experienced perforation or laceration of the IVC during filter retrieval, theoretically minor IVC perforations could potentially be managed with observation and gentle balloon tamponade because the IVC is a relatively lowpressure system. Frank extravasation of contrast suggests either a longitudinal or circumferential laceration, which would potentially require open repair or treatment with a covered endograft.
FILTER SAFETY
Overall, IVC filters do what they were designed to
do—mechanical prevention of PE by trapping thrombus.
Data demonstrate that in-hospital recurrent PE is
less and survival greater with the use of anticoagulation
and filters than with anticoagulation alone.21 The judicious
use of filters appears safe in appropriately selected
patients, and long-term data suggest that filters do
provide outcome benefits with regard to prevention of
PE at a cost of a slightly increased risk for developing
lower extremity swelling and recurrent deep vein
thrombosis.10,22,23
CONCLUSION
With the development of optional IVC filters, their
use has dramatically increased during the past decade
despite the absence of level I data supporting its use.
For this reason, the most important aspect of using an
IVC filter is to ensure that there is an appropriate and
good indication for its use. The operator should be
very familiar with the device and the anatomy in which
the filter is to be placed, knowledgeable and aware of
normal variants, and use thoughtful and meticulous
procedural planning and techniques to optimize the
efficacy of the IVC filter, while minimizing the chance
for a misadventure during device deployment.
Acknowledgements: The authors thank Leanne Dore Lessley, RT (R) VI, for her artistic contribution of Figures 2 and 4, and Lauren J. Germain for her help in preparing this article.
Ulku Cenk Turba, MD, is Assistant Professor of Interventional Radiology in the Department of Radiology at the University of Virginia in Charlottesville, Virginia. He has disclosed that he holds no financial interest in any product or manufacturer mentioned herein. Dr. Turba may be reached at turba@virginia.edu.
Saher S. Sabri, MD, is Assistant Professor of Interventional Radiology in the Department of Radiology at the University of Virginia in Charlottesville, Virginia. He has disclosed that he holds no financial interest in any product or manufacturer mentioned herein. Dr. Sabri may be reached at (434) 924-9401; ss2bp@virginia.edu.
Wael E.A. Saad, MD, is Associate Professor in the Department of Radiology at the University of Virginia in Charlottesville, Virginia. He has disclosed that he holds no financial interest in any product or manufacturer mentioned herein. Dr. Saad may be reached ws6r@virginia.edu.
Auh Whan Park, MD, is Associate Professor in the Department of Radiology at the University of Virginia in Charlottesville, Virginia. He has disclosed that he holds no financial interest in any product or manufacturer mentioned herein. Dr. Park may be reached at ap7we@virginia.edu.
John F. Angle, MD, is Professor and Division Chief of
Interventional Radiology at the University of Virginia in Charlottesville,
Virginia. He has disclosed that he receives research support from Terumo Interventional Systems. Dr. Angle may be reached at jfa3h@virginia.edu.
Alan H. Matsumoto, MD, is Professor of Radiology and Chair of the Department of Radiology at the University of Virginia in Charlottesville, Virginia. He has disclosed that he receives grant/research funding from Cook Medical, Medtronic, W. L. Gore & Associates, Endologix, Talecris Biotherapeutics, InSightec, and NIH. He is the DSMB Chair for Bolton Medical's TEVAR trial and serves on the advisory board for Boston Scientific, Siemens Medical, and Bard Peripheral Vascular. He has also disclosed that he receives honoraria from Bard Peripheral Vascular and AngioScore. Dr. Matsumoto may be reached at (434) 924 9279; ahm4d@virginia.edu.