The VIBRANT Trial
Comparing bare-nitinol stents to stent grafts in long superficial femoral artery lesions in this unique trial.
By Gary M. Ansel, MD, FACC; Patrick J. Geraghty, MD, FACS; Mark Mewissen, MD; and Michael R. Jaff, DO
Management of patients with femoropopliteal peripheral artery disease must be directed so that maximum clinical efficacy is obtained with acceptable short- and long-term safety. The adoption of endovascular therapy for the superficial femoral artery (SFA), often without supporting data, has inspired considerable debate. Unfortunately, many of the issues raised in these debates are difficult to resolve due to disparate reporting standards in the literature. Results of percutaneous transluminal angioplasty (PTA) of moderately complex SFA lesions have demonstrated poor sustained patency.1 Both adjunctive nitinol stenting and stent grafting have published data demonstrating superiority to PTA in moderately complex lesions at 1-year follow-up.2-4 Although surgical venous bypass can achieve similar or better patency, use of bare and expanded polytetrafluoroethylene-lined stents offers a significantly less invasive approach to restoring artery flow. Recently, a single-center, randomized trial comparison of stent grafting to surgical prosthetic bypass demonstrated equivalence in complex SFA lesions.5
THE VIBRANT TRIAL
The VIBRANT (Viabahn Versus Bare Nitinol Stent in the
Treatment of Long Lesion [≥ 8 cm] Superficial Femoral
Artery Occlusive Disease) trial was developed to compare
bare-nitinol stenting in complex real-world patients to the
Viabahn stent graft (W. L. Gore & Associates, Flagstaff, AZ).
This investigation has several unique and independent features:
(1) evaluate SFA treatment with femoropopliteal
lesion lengths > 8 cm, allowing for stent overlap; (2) evaluate
independent core lab adjudication of all major outcomes;
(3) evaluate prospective correlation of various duplex ultrasound
systolic velocity ratios with recurrence of clinical
symptoms; (4) evaluate potential predictors of patency with
a focus on longer-term follow-up (3 years); and (5) evaluate
the significance of device fractures on clinical and restenosis
endpoints.
Study Design
VIBRANT is a prospective, multicenter, randomized trial
comparing the nonheparin-bonded Viabahn stent graft to
bare-nitinol stents. Eighteen centers from around the
United States representing all three disciplines that practice
endovascular treatment were included. VasCore
(Massachusetts General Hospital, Boston, MA), the vascular
ultrasound core laboratory, performed independent assessments
of all duplex ultrasound and radiographic images for
evidence of restenosis and stent fracture, respectively. The
study hypothesis states that in comparison to the use of
bare-nitinol stents in treating chronic SFA long lesions (≥ 8
cm), stenoses, and occlusions, the use of the Gore Viabahn
endoprosthesis will result in greater midterm (24 months)
and long-term (36 months) patency of the treated arterial
lesion.
Study Devices
The Gore Viabahn endoprosthesis (US Food and Drug
Administration approved for SFA indication on June 14,
2005) is constructed with a very flexible but durable, reinforced,
biocompatible, expanded polytetrafluoroethylene
liner attached to an external nitinol stent structure and is
indicated for improving blood flow in patients with symptomatic
peripheral artery disease due to SFA lesions with reference
vessel diameters ranging from 4 to 7.5 mm (Figure 1). Bare-nitinol stents, as determined by the institutional standard
of care when treating SFA occlusive disease, were also
used. Permitted device diameters for the study were 6, 7,
and 8 mm only.
Study Guidelines
The study population includes 150 patients randomized
into the test and control groups. Inclusion criteria included
Rutherford class 1 to 4. De novo stenosis or post-PTA
restenosis (> 50% by visual estimate) or occlusion of the
native SFA ≥ 8 cm (TASC lesion types C and D) in length
were located in the region beginning 1 cm below the origin
of the profunda femoris artery and ending 5 cm above the
radiographic knee joint. Any previous PTA, if on the target
lesion, must have been performed before the study procedure
by at least 6 months. Acceptable reference vessel
diameters were 4.8 to 7.5 mm in the treatment segments
within the SFA and proximal popliteal artery.
Stenoses were predilated with angioplasty balloons sized to a diameter slightly smaller than the native vessel. Investigators were directed to completely cover the treated site plus 5 mm on both ends with either the stent or stent graft. Postballoon dilation was to be performed within the stent graft or stent to the diameter of the native vessel. If the patient was not currently taking clopidogrel, a 600-mg oral loading dose was administered immediately after successful intervention; a 75-mg oral dose of clopidogrel was continued once daily for a minimum of 6 months after the procedure. An oral dose of aspirin (81 or 325 mg) was continued once daily through study follow-up. Flat plate x-ray images for determination of stent fracture were made at multiple angles with patient extremities straight and flexion.
Study Endpoints
Primary endpoints included 3-year primary patency
(duplex peak systolic velocity ratio ≤ 2 based on core lab
determination) for efficacy, with a 30-day composite safety
endpoint inclusive of major complications. Secondary endpoints
included primary assisted patency, secondary patency,
technical success, target vessel revascularization, target
lesion revascularization, clinical success, change in anklebrachial
index from baseline, alternate peak systolic velocity
ratios of 2.5 and 3, and stent fracture on plain radiographs.
Quality of life and standardized claudication questionnaires
were administered at baseline, postprocedure, and annual
follow-up.
Study Status
To date, all patients have been enrolled and followed for
1 year. Interim results were presented at the 2009 VIVA
meeting in Las Vegas, Nevada. Longer-term follow-up continues on the entire cohort, and the primary endpoint will
be adjudicated in 2011.
Since the start of the VIBRANT trial, the Viabahn endoprosthesis has undergone several iterative modifications. The device now has a lower delivery profile, heparin bonding, and, most recently, the proximal edge has been contoured. The VIPER registry trial, discussed by Richard R. Saxon, MD, on page 66 in this issue, is underway to evaluate patency in a similar population as VIBRANT.
Gary M. Ansel, MD, FACC, is Director of the Critical Limb Care Center at Riverside Methodist Hospital in Columbus, Ohio. He has disclosed that he receives research funding from Abbott Vascular, Boston Scientific Corporation, Cook Medical, and W. L. Gore & Associates and that he serves on the advisory board for Boston Scientific Corporation, Cordis Corporation, Flexible Stenting Solutions, and W. L. Gore & Associates. He has also disclosed that he serves on the speakers bureaus for Abbott Vascular and W. L. Gore & Associates, that he is a speaker for Cordis Corporation, and that he holds product royalties in Cook Medical. Dr. Ansel may be reached at (614) 262-6772; gansel@mocvc.com.
Patrick J. Geraghty, MD, FACS, is Associate Professor of Surgery and Radiology at Washington University in St. Louis, Department of Surgery, Section of Vascular Surgery in St. Louis, Missouri. He has disclosed that he receives grant/research funding from Cook Medical and is a paid consultant to W. L. Gore & Associates and Pulse Therapeutics. Dr. Geraghty may be reached at (314) 362-6490; geraghtyp@wustl.edu.
Mark Mewissen, MD, is Director of the Vascular Center at St. Luke's Medical Center in Milwaukee, Wisconsin. He has disclosed that he holds no financial interest in any product or manufacturer mentioned herein. Dr. Mewissen may be reached at (414) 385-2429; mark.mewissen@aurora.org.
Michael R. Jaff, DO, is Associate Professor of Medicine, Harvard Medical School, and Medical Director, Vascular Center, Massachusetts General Hospital in Boston. He has disclosed that he holds no financial interest in any product or manufacturer mentioned herein. Dr. Jaff may be reached at (617) 726-3784; mjaff@partners.org.