Critical Decision Making in Carotid Embolic Protection
Carotid stenting in a patient with stenosis of the left external and internal carotid arteries.
By Jennifer Franke, MD; Marius Hornung; Nina Wunderlich, MD; and Horst Sievert, MD, FESC, FACC, FSCAI
In September 2010, a 72-year-old man was admitted to the neurological department of his nearest hospital due to a transient ischemic attack with sudden weakness of the right arm and leg that continued for approximately 10 minutes. Cranial magnetic resonance imaging ruled out cerebral lesions. His risk factors included hypertension (adequately treated), hyperlipidemia, and peripheral arterial occlusive disease. Initial examinations included a duplex scan of the supra-aortic arteries, a transcranial Doppler examination, transthoracic and transesophageal echocardiography, and a 24-hour electrocardiogram. Diagnostics revealed a high-grade eccentric stenosis of the left internal carotid artery (ICA) and the left external carotid artery (ECA).
After informing the patient about different treatment modalities, clopidogrel and statins were added to his medication, and he was referred to our center for carotid stenting of the symptomatic left ICA. Angiography demonstrated a short, eccentric, high-grade stenosis of the left ICA, an ostial subtotal occlusion of the left ECA with slow flow, and a comparably large-lumen superior thyroid artery. As measured by angiography, the common carotid artery (CCA) measured 7 mm in diameter, and the ICA was 5 mm in diameter. The intracranial runs revealed adequate flow in both the anterior and middle cerebral artery.
DECISION POINT 1
Considering the patient suffered a transient ischemic attack
1 week before, would a proximal embolic protection device or
a distal filter device be best in this situation?
Because the procedure was performed shortly after a cerebrovascular
event, embolic complications were more likely
to occur due to vulnerable plaques.1-3 By providing complete
protection through retrograde flow before lesion manipulation,
the chance of microembolization of very small particles
(< 80 µm in diameter) can theoretically be reduced by using
a proximal occlusion device. These systems consist of a long
introducer sheath with a balloon that is inflated in the CCA.
A second balloon inflated in the ECA ensures the total
blockade of the antegrade blood flow in the ICA. The proximal
protection systems facilitate the cerebral vascular connections
of the circle of Willis. After the occlusion of the
CCA and ECA, the collateral flow through the circle of Willis
creates so-called back pressure, which prevents antegrade
flow in the ICA.
Although large, randomized studies comparing the clinical benefit of the different approaches are pending, it is conceivable that a proximal occlusion device could be beneficial in vulnerable lesions with fresh thrombus compared to a device that is placed distally and therefore requires unprotected crossing of the lesion. With the Mo.Ma system (Medtronic Invatec, Frauenfeld, Switzerland), the occlusion balloons of the ECA and CCA are mounted on the guiding catheter in a fixed distance. Therefore, in patients with significant stenosis of the ECA, positioning of the distal balloon is either cumbersome or, in some cases, not possible.4-5 In the situation of a total occlusion of the ECA, the sole occlusion of the CCA balloon may not establish complete blockage of antegrade flow. The total occlusion may lie distal to the ostium of the most proximal branches of the ECA—the superior thyroid artery and lingual artery.
In our patient, the ECA was subtotally occluded at its ostium, with only slow flow through the external carotid lesion shown in angiographic imaging. In addition, the patient had an anatomical variation—the superior thyroid artery originated from the CCA below the carotid bifurcation and not as a side branch of the ECA distal to the subtotal lesion (Figure 1A). In this quite specific anatomy, the use of the Mo.Ma system was limited due to the previously mentioned specifications of the device. Unlike with the Mo.Ma device, the proximal and distal balloons of the Gore Flow Reversal system (Gore & Associates, Flagstaff, AZ) are mounted on two separate components. The distal balloon is mounted on the balloon wire, and the proximal balloon is located at the distal end of the introducer sheath. On the basis of these small advantages, the Gore Flow Reversal system was a viable alternative in our patient's anatomy because it allowed the operator to position the proximal occlusion balloon at any distance to the bifurcation. However, after positioning the proximal protection device and occluding the balloon in the CCA, a column of contrast remained in the CCA, and the contrast located in the ICA was washed out slowly. In this situation, forward flow to the intracranial arteries was still maintained via the superior thyroid artery, partially contributed to by the subtotally occluded ECA. Thus, adequate adequate cerebral protection was not achieved (Figure 1B).
DECISION POINT 2
What should you do when antegrade
flow cannot be adequately
blocked by proximal occlusion?
Some options are available to
prevent antegrade flow in this situation.
One basic concept is to
increase the amount of back pressure
applied to the cerebral circulation.
By opening the proximal stopcock
of the arterial sheath and letting
it bleed back, or alternatively,
by attaching the external filter of
the device and connecting the system
to a sheath placed in a femoral
vein (producing an arteriovenous
shunt), retrograde flow can be
enhanced. Due to the patient's specific
anatomical situation with the
superior thyroid artery originating
from the CCA below the carotid
bifurcation, blockage of the superior thyroid artery with a
balloon would prevent antegrade flow via this route, yet
would not eliminate the chance of distal embolization via
the ECA. Performing manual aspirations with a syringe
between the steps of the procedure would additionally prevent
distal embolization but would not sufficiently eliminate
antegrade flow in between aspirations.
A further option would be to place a filter embolic protection device in the distal ICA and therefore achieve additional distal protection by filtering the remaining antegrade blood flow. In this particular patient, as complete retrograde flow could not be established even after connecting the system to the femoral vein, the operator decided to place a Gore Embolic Filter device (Gore &Associates) adjunctive to the proximal protection device (Figure 2). A feature of the Gore Embolic Filter is its comparably long circumferential frame at the proximal end of the filter, which provides improved alignment of the filter basket to the vessel wall. As a result of filter placement, embolic protection was additionally given through the filtration of blood that continued to flow distally despite proximal protection.
DECISION POINT 3
Which stent design is preferred in symptomatic carotid
lesions?
The clinical impact of open-cell design versus closed-cell
design, or perhaps more importantly the impact of cell size
and pore size, is still unclear. Insufficient wall coverage of a stent may lead to protruding atherosclerotic debris through
the stent struts. Therefore, in lesions with high emboligenic
potential, as presumed in this patient, closed-cell stents may
provide better scaffolding.6-8 Additionally, choosing a
carotid stent that is slightly larger than the vessel diameter
itself (oversizing), creates an artificially smaller pore size
between the stent struts. In this case, an 8-mm X 3-cm Xact
stent (Abbott Vascular, Santa Clara, CA) was implanted and
postdilated with a 5-mm balloon. Angiographic results were
satisfactory (Figure 1C and 1D). Throughout the procedure
and during the postprocedural period, the patient showed
no neurological symptoms or hemodynamic complications.
He was discharged 2 days after the procedure.
CONCLUSION
When using the Gore Flow Reversal system, it is mandatory
to verify complete blockage of antegrade flow in the vessel
and the achievement of blood flow reversal in the ICA
before continuing with the next step of the intervention. If
blood flow in the ICA merely stagnates, cerebral protection
is not sufficient. In this situation, a short interruption of
back pressure during the procedure could allow debris
accrued in this area to flow cranially and possibly cause a
severe stroke. In the case of the Mo.Ma device, the operator
needs to verify that there is no flow in the ICA after inflation
of the clamping balloons in the ECA and CCA. Furthermore, performing the intervention stepwise and aspirating in
between the procedural steps may provide additional safety.
In general, correct selection of embolic protection based
on clinical aspects as well as on radiological features of the
lesion is essential and greatly influences the outcome of
carotid stenting. Particularly in challenging anatomical situations,
such as in the case presented here, specific knowledge
of device-related advantages and potential drawbacks of
different concepts of embolic protection is required to provide
patients with the optimal treatment modality.
Jennifer Franke, MD, is a fellow in cardiology and performed her doctorate thesis at the CardioVascular Center Frankfurt, Germany. She has disclosed that she holds no financial interest in any product or manufacturer mentioned herein.
Marius Hornung, is a doctoral candidate at the CardioVascular Center Frankfurt in Frankfurt, Germany. He has disclosed that he holds no financial interest in any product or manufacturer mentioned herein.
Nina Wunderlich, MD, is Codirector of the CardioVascular Center Frankfurt in Frankfurt, Germany. She has disclosed that she holds no financial interest in any product or manufacturer mentioned herein.
Horst Sievert, MD, FESC, FACC, FSCAI, is Director of the CardioVascular Center Frankfurt in Frankfurt, Germany. He has disclosed that he has received consulting fees, travel expenses, or honoraria for clinical trials from Abbott Vascular, Boston Scientific Corporation, Cordis Corporation, EndoTex Interventional Systems, ev3, Inc., Gore & Associates, Guidant Corporation, Lumen Biomedical, and Medtronic Invatec. Dr. Sievert may be reached at horstsievertmd@aol.com.