Dramatic improvements have been made in the care of patients harboring vascular disease over the past 2 decades. Much of this progress has been made on the back of new device design. In 2008, the American Association for the Surgery of Trauma published results on emerging trends in the management of blunt aortic injury (BAI) and stated that, “There is a major and urgent need for improvement of the available endovascular devices.”1 Industry responded to this call for better device design with improvements that have finally arrived. In 2010, I was invited by Cook Medical to serve as Principal Investigator for TRANSFIX, the national multicenter clinical trial evaluating the Zenith TX2 low-profile endovascular graft (now called Zenith Alpha Thoracic) for the management of patients presenting with BAI. The following is a description of a few cases using this device to manage severely injured patients with aortic injury.

Figure 1. The precurved inner cannula.

DISCUSSION

The Zenith Alpha Thoracic device offers what amounts to a great breakthrough in managing patients with BAI. The low-profile, hydrophilic, braided sheath delivery system; precurved inner cannula (Figure 1); and nitinol-based stent design provide for unparalleled opportunity to treat a wide variety of patients. With the lowest treatable aortic diameter (15 mm), lowest arch radius indication (20 mm), and smallest-diameter delivery system (16 F), more patients can be treated with this newer-generation device. A comparison of Zenith Alpha Thoracic with its predecessor, Zenith TX2, is depicted in Table 1.

TRANSFIX TRIAL DESIGN AND SHORT-TERM RESULTS

Fifty patients were enrolled into the prospective, nonrandomized TRANSFIX trial between January 2013 and May 2014. Patients in the trial will be followed through 5 years. The primary safety endpoint is 30-day mortality, and the primary efficacy endpoint is 30-day device success. As presented at the 2014 annual meeting of the Society for Vascular Surgery, technical success was achieved in all patients (100%), and there were no intraoperative mortalities. Short-term results indicate that the Zenith Alpha Thoracic device appears safe and effective for the management of patients with BAI. As of October 2015, the Zenith Alpha Thoracic device has been approved for use by the US Food and Drug Administration.

Other than access-related complications, the most feared complication of thoracic endovascular aortic repair for BAI is either stroke or paraplegia. Modern workup includes magnetic resonance (MR) imaging of the brain or spinal cord, respectively. In the past, the presence of ferrous stent graft designs in the thoracic aorta was a contraindication to MR imaging in these scenarios. The Zenith Alpha Thoracic device has improved compatibility with MR imaging, which allows for alternative imaging in challenging clinical scenarios.

CASE STUDY

Figure 2

Figure 3

Figure 4

Figure 5

Figure 6

Figure 7

CONCLUSION

Zenith Alpha Thoracic represents a powerful tool in our armamentarium for managing aortic pathology. The management of BAI has become a percutaneous, semielective procedure that can be performed in under an hour. Thanks to better device design that includes a smaller, precurved delivery system and a nitinol frame, more patients with BAIs are candidates for this minimally invasive technology.

1. Demetriades D, Velmahos GC, Scalea TM, et al. Operative repair or endovascular stent graft in blunt traumatic thoracic aortic injuries: results of an American Association for the Surgery of Trauma Multicenter Study. J Trauma. 2008;64:561-570; discussion 570-571.

Benjamin W. Starnes, MD, FACS
Professor and Chief
Division of Vascular Surgery
Vice Chair
Department of Surgery
University of Washington
Seattle, Washington
starnes@uw.edu
Disclosures: University of Washington Medical Center
receives payment for service performed by Dr. Starnes
on behalf of Cook Medical.